Cardiac resynchronization therapy
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Bhaskar Purushottam, M.D. [2] Nehal Eid, M.D.[3]
Synonyms and keywords: CRT
Overview
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Bhaskar Purushottam, M.D. [2]
Overview
Cardiac resynchronization therapy (CRT) is an evidence based device treatment for congestive heart failure. CRT is indicated in those patients with symptomatic congestive heart failure despite optimal medical therapy who have a reduced left ventricular ejection fraction(an LVEF ≤ 35%), and a wide QRS (≥ 0.12 sec). It involves timed atrioventricular and biventricular pacing, which can improve left ventricular function, heart failure symptoms and may be associated with a reduction in mortality.
References
Indications
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief: Sara Zand, M.D.[2] Bhaskar Purushottam, M.D. [3], Hardik Patel, M.D.
Overview
Cardiac resynchronization therapy (CRT) with or without an implantable cardiac defibrillator (ICD) is indicated in patients who have an LVEF less than or equal to 35%, left bundle branch block (LBBB) with a QRS duration greater than or equal to 0.15 seconds, and normal sinus rhythm, for the treatment of NYHA functional Class III or ambulatory Class IV heart failure symptoms in patients whose medical therapy has been optimized.
Indications
2021 ESC Guideline for Cardiac Resynchronization Therapy implantation
Abbreviations:
AF: Atrial fibrillation;
A-V: Atrio-ventricular;
CRT: Cardiac resynchronization therapy ;
HFrEF: Heart failure with reduced ejection fraction;
ICD: Implantable cardioverter-defibrillato;
LBBB:Left bundle branch block;
LVEF: Left ventricular ejection fraction;
NYHA:New York Heart Association;
RV: = Right ventricular
| Recommendations for cardiac resynchronization therapy implantation in patients with heart failure |
| (Class I, Level of Evidence A): |
|
❑ CRT is recommended for symptomatic patients with HF in sinus rhythm with a QRS duration ≥150 ms and LBBB QRS morphology and with LVEF ≤35%
despite optimal medical therapy in order to improve symptoms and reduce morbidity and mortality |
| (Class IIa, Level of Evidence B): |
|
❑ CRT should be considered for symptomatic patients with HF in sinus rhythm with a QRS duration ≥150 ms and non-LBBB QRS morphology and
with LVEF ≤35% despite OMT in order to improve symptoms and reduce morbidity and mortality |
| (Class IIb, Level of Evidence B) : |
|
❑ CRT may be considered for symptomatic patients with HF in sinus rhythm with a QRS duration of 130-149 ms and non-LBBB QRS morphology and with LVEF ≤35% despite optimal medical therapy in order to improve symptoms and reduce morbidity and mortality |
| (Class III, Level of Evidence A) : |
|
❑CRT is not recommended in patients with a QRS duration <130 ms who do not have an indication for pacing due to high degree AV block |
| The above table adopted from 2021 ESC Guideline |
|---|
- CRT reduces morbidity and mortality.[2]
- CRT improves cardiac function, and enhances quality of life.
- Several characteristics predictors of improvement in morbidity and mortality including the extent of reverse remodelling as the most important mechanisms of action of CRT.
- Patients with HFrEF and ischemic etiology have less improvement in LV function due to myocardial scar tissue, which is less likely to undergo favorable remodelling.
- Women may be more likely to respond to CRT than men, possibly due to smaller body and heart size.[2]
- QRS duration predicts CRT response.[3]
- QRS morphology is related to a beneficial response to CRT.
- Patients with left bundle branch block (LBBB) morphology are more likely to respond favorably to CRT, whereas there is less certainty about patients with non-LBBB morphology.
- Patients with LBBB morphology often have wider QRS durations, and there is a current debate about whether QRS durations or QRS morphology is the
the main predictor of a beneficial response to CRT.[4]
- there is little evidence to suggest that QRS morphology or etiology of disease influence the effect of CRT on morbidity or mortality.
- Implantation of CRT is not recommended if QRS duration is <130 ms.
- If a patient is selected to receive an ICD and is in sinus rhythm, with a LBBB, CRT-D should be considered if the QRS is between 130 and 149 ms and is recommended if QRS is ≥ 150 ms.
- When LVEF is reduced, RV pacing may exacerbate cardiac dyssynchrony.
- This can be prevented by CRT, which might improve patient outcomes.
- CRT rather than RV pacing is recommended for patients with HFrEF regardless of NYHA class who have an indication for ventricular pacing in order to reduce morbidity, although no clear effect on mortality was observed.
- In patients with HFrEF who have received a conventional pacemaker or an ICD and subsequently develop worsening HF with a high proportion of RV pacing, CRT implantation is recommended.[5]
- CRT is superior to RV pacing in patients undergoing atrioventricular (AV) node ablation in AF patients.[6]
- In patients with AF, CRT-D compared with ICD, was not benefit and less than half of patients had >90% biventricular capture.
- CRT in patients with AF may be an option in selected patient with QRS≥ 150 ms, high biventricular pacing.[7]
- When biventricular capture is <98%, it reflects a loss of resynchronization and poor prognosis in CRT.
- Patients with extensive myocardial scar will have less improvement in LV function with CRT.
- Pacing thresholds are higher in scarred myocardium and, if possible, lead placement should avoid such regions.
- Although patients with extensive scarring have an intrinsically worse prognosis, there is little evidence that they obtain less prognostic benefit from CRT.
- Optimization of AV intervals or interventricular delay intervals (VV intervals) after implantation by using echo– or electrocardiographic criteria or BP response may be considered for patients who have had no response to CRT.[4]
References
- ↑ McDonagh TA, Metra M, Adamo M, Gardner RS, Baumbach A, Böhm M, Burri H, Butler J, Čelutkienė J, Chioncel O, Cleland J, Coats A, Crespo-Leiro MG, Farmakis D, Gilard M, Heymans S, Hoes AW, Jaarsma T, Jankowska EA, Lainscak M, Lam C, Lyon AR, McMurray J, Mebazaa A, Mindham R, Muneretto C, Francesco Piepoli M, Price S, Rosano G, Ruschitzka F, Kathrine Skibelund A (September 2021). “2021 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure”. Eur Heart J. 42 (36): 3599–3726. doi:10.1093/eurheartj/ehab368. PMID 34447992 Check
|pmid=value (help). Vancouver style error: initials (help) - ↑ 2.0 2.1 Woods B, Hawkins N, Mealing S, Sutton A, Abraham WT, Beshai JF, Klein H, Sculpher M, Plummer CJ, Cowie MR (November 2015). “Individual patient data network meta-analysis of mortality effects of implantable cardiac devices”. Heart. 101 (22): 1800–6. doi:10.1136/heartjnl-2015-307634. PMC 4680159. PMID 26269413.
- ↑ Moss AJ, Hall WJ, Cannom DS, Klein H, Brown MW, Daubert JP, Estes NA, Foster E, Greenberg H, Higgins SL, Pfeffer MA, Solomon SD, Wilber D, Zareba W (October 2009). “Cardiac-resynchronization therapy for the prevention of heart-failure events”. N Engl J Med. 361 (14): 1329–38. doi:10.1056/NEJMoa0906431. PMID 19723701.
- ↑ 4.0 4.1 Cleland JG, Abraham WT, Linde C, Gold MR, Young JB, Claude Daubert J, Sherfesee L, Wells GA, Tang AS (December 2013). “An individual patient meta-analysis of five randomized trials assessing the effects of cardiac resynchronization therapy on morbidity and mortality in patients with symptomatic heart failure”. Eur Heart J. 34 (46): 3547–56. doi:10.1093/eurheartj/eht290. PMC 3855551. PMID 23900696.
- ↑ Curtis AB, Worley SJ, Adamson PB, Chung ES, Niazi I, Sherfesee L, Shinn T, Sutton MS (April 2013). “Biventricular pacing for atrioventricular block and systolic dysfunction”. N Engl J Med. 368 (17): 1585–93. doi:10.1056/NEJMoa1210356. PMID 23614585.
- ↑ Brignole M, Botto G, Mont L, Iacopino S, De Marchi G, Oddone D, Luzi M, Tolosana JM, Navazio A, Menozzi C (October 2011). “Cardiac resynchronization therapy in patients undergoing atrioventricular junction ablation for permanent atrial fibrillation: a randomized trial”. Eur Heart J. 32 (19): 2420–9. doi:10.1093/eurheartj/ehr162. PMID 21606084.
- ↑ Leclercq C, Walker S, Linde C, Clementy J, Marshall AJ, Ritter P, Djiane P, Mabo P, Levy T, Gadler F, Bailleul C, Daubert JC (November 2002). “Comparative effects of permanent biventricular and right-univentricular pacing in heart failure patients with chronic atrial fibrillation”. Eur Heart J. 23 (22): 1780–7. doi:10.1053/euhj.2002.3232. PMID 12419298.
Contraindications
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Bhaskar Purushottam, M.D. [2] Synonyms and Keywords: CRT
Overview
CRT is contraindicated in patients who are asymptomatic or who have a limited life expectancy.
Contraindications
- Patients who are asymptomatic with a reduced LVEF who have no other indications for pacing.
- Patients who have a limited life expectancy due to a non-cardiac condition.
- Patients who have a limited functional capacity due to a chronic non-cardiac condition.
ACC / AHA Guidelines – Recommendations for Cardiac Resynchronization Therapy in Patients with Severe Systolic Heart Failure (DO NOT EDIT)[1]
| “ |
1) CRT is not indicated for asymptomatic patients with reduced LVEF in the absence of other indications for pacing. (Level of Evidence: B) 2) CRT is not indicated for patients whose functional status and life expectancy are limited predominantly by chronic noncardiac conditions.(Level of Evidence: C)
|
” |
References
- ↑ Epstein AE, DiMarco JP, Ellenbogen KA, Estes NA, Freedman RA, Gettes LS, Gillinov AM, Gregoratos G, Hammill SC, Hayes DL, Hlatky MA, Newby LK, Page RL, Schoenfeld MH, Silka MJ, Stevenson LW, Sweeney MO, Smith SC, Jacobs AK, Adams CD, Anderson JL, Buller CE, Creager MA, Ettinger SM, Faxon DP, Halperin JL, Hiratzka LF, Hunt SA, Krumholz HM, Kushner FG, Lytle BW, Nishimura RA, Ornato JP, Page RL, Riegel B, Tarkington LG, Yancy CW (2008). “ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons”. Circulation. 117 (21): e350–408. doi:10.1161/CIRCUALTIONAHA.108.189742. PMID 18483207. Retrieved 2011-01-15. Unknown parameter
|month=ignored (help)
Pathophysiologic Basis For CRT
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Bhaskar Purushottam, M.D. [2]
Overview
Left ventricular systolic dysfunction is associated with electromechanical dyssynchrony which may be in the form of atrioventricular, interventricular, intraventricular and intramural delay. The impaired electromechanical coupling in turn diminishes the left ventricular systolic function.
Pathophysiologic Basis For CRT
Left ventricular systolic dysfunction is often accompanied by impaired electromechanical coupling, which may further diminish the left ventricular systolic function. The types of electromechanical dyssynchrony are atrioventricular, interventricular, intraventricular and intramural delay[1]. Atrioventricular dyssynchrony results in a late diastolic ventriculoatrial gradient and so called “pre-systolic” mitral regurgitation. Interventricular dyssynchrony is the time delay between the contraction of the left and right ventricles and this is calculated by measuring the difference in the time of onset of systolic flow in the aortic and pulmonic valve. A time difference greater than or equal to 40 milliseconds is indicative of interventricular dyssynchrony. There are several echocardiographic techniques to measure intraventricular dyssynchrony, which include M mode echocardiography, tissue Doppler imaging, tissue strain, strain rate analysis and speckle tracking echocardiography. Intramural dyssynchrony is the dyssynchrony within the myocardial wall and it has been measured using speckle tracking echocardiography[2]. However, the most common abnormalities are prolonged atrio-ventricular and ventricular conduction, which causes regional mechanical delay within the left ventricle.
This mechanical delay is responsible for ventricular dyssynchrony, which can result in the following hemodynamic abnormalities[3]:
- Reduced left ventricular systolic function, decreased dP/dT and cardiac output
- Delayed contraction of lateral and posterior left ventriclular wall with abnormal septal motion
- Increased myocardial energy expenditure
- Adverse remodeling with increased left ventricular dilatation left ventricular end systolic volume
- Functional mitral regurgitation
- Delayed mitral valve opening
- Reduced left ventricular filling time with increased left atrial pressures
- Distorted mitral valve annulus
- Delayed aortic valve opening and closure with reduced systolic ejection time
References
- ↑ Auricchio A, Abraham WT (2004). “Cardiac resynchronization therapy: current state of the art: cost versus benefit”. Circulation. 109 (3): 300–7. doi:10.1161/01.CIR.0000115583.20268.E1. PMID 14744954.
- ↑ Bank AJ, Kaufman CL, Burns KV, Parah JS, Johnson L, Kelly AS; et al. (2010). “Intramural dyssynchrony and response to cardiac resynchronization therapy in patients with and without previous right ventricular pacing”. Eur J Heart Fail. 12 (12): 1317–24. doi:10.1093/eurjhf/hfq162. PMID 20864481.
- ↑ Ho JK, Mahajan A (2010). “Cardiac resynchronization therapy for treatment of heart failure”. Anesth Analg. 111 (6): 1353–61. doi:10.1213/ANE.0b013e3181fa3408. PMID 21059745.
Procedure
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Bhaskar Purushottam, M.D. [2]
Overview
CRT requires implantation of atrial and right ventricular leads, as well as a third lead in the coronary sinus and the lateral or posterior branch to stimulate the left ventricle.
Procedure
CRT involves a procedure similar to that of a pacemaker placement. In addition to that of the routine implantation of the atrial and right ventricular lead, a third lead is introduced into the coronary sinus and the lateral or posterior branch is accessed for stimulation of the left ventricle. Previously, an epicardial left ventricular lead was implanted after a limited lateral throacotomy. Such epicardial lead implantation is associated with high capture thresholds, suboptimal position for resynchronization, a far more invasive procedure, risk of general anaesthesia and standard complications associated with thoracotomy. However, this approach may be used if the coronary sinus or the appropriate branch cannot be accessed due to anatomical variations, vein stenosis, coronary sinus injury, tortuosity of the coronary sinus and distortion of the ostium.
References
Outcomes and Prognosis
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Bhaskar Purushottam, M.D. [2]
Overview
Approximately 30% of patients undergoing CRT will not sustain clinical or functional improvements for a wide variety of reasons.
Non-responders
30% of the CRT recipients are considered “non-responders”. A patient is considered a “non-responder” if there is no significant clinical or functional improvement following CRT. There are multiple explanations as to why some patients are non-responders.
- Not all patients with QRS duration greater than or equal to 0.12 seconds have mechanical dyssynchrony. Unfortunately, the PROSPECT[1] trial which set out to examine the various echocardiographic parameters to predict CRT response was not successful. Some of the major limitations in the study were the technical difficulties in obtaining the dyssynchrony parameters and the discrepancies among the different centers.
- The leads may have been placed in regions of the left ventricle which are not dyssynchronous
- The leads may have been placed in regions with fibrosis or viable myocardium.
- Anterior left ventricular lead placement has been associated with worsening hemodynamics.
- Non-responsiveness can occur due to high left ventricular capture thresholds
- Lead dislodgement
- Long atrioventricular delay
- Atrial tachyarrhythmias with rapid ventricular response
- Frequent premature ventricular contractions.
- Lack of optimal atrioventricular and ventricular to ventricular (i.e., right ventricle to left ventricle) timing can result in non-responsiveness.
References
- ↑ Chung ES, Leon AR, Tavazzi L, Sun JP, Nihoyannopoulos P, Merlino J; et al. (2008). “Results of the Predictors of Response to CRT (PROSPECT) trial”. Circulation. 117 (20): 2608–16. doi:10.1161/CIRCULATIONAHA.107.743120. PMID 18458170.
Device management
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Nehal Eid, M.D.[2]
Device management
Post-implantation management of cardiac resynchronization therapy (CRT) focuses on maintaining effective resynchronization, detecting device- and heart-failure–related deterioration, managing apparent non-response, and planning generator replacement, imaging, infection care, and goals-of-care discussions. Initial implantation technique, primary CRT indications, and primary conduction-system pacing strategies are addressed elsewhere.
Routine device surveillance and programming
| Area | Device-management approach |
|---|---|
| Effective biventricular pacing | Aim for ≥98% effective biventricular (BiV) pacing. Device-reported pacing percentages may overestimate true resynchronization when triggered LV pacing, fusion, or pseudofusion is present; review stored electrograms and consider ambulatory ECG monitoring when effective pacing is uncertain.[1][2] |
| Causes of reduced effective pacing | Identify and correct atrial tachyarrhythmias, premature ventricular contractions, inappropriate AV timing, intermittent LV capture, and lead-related problems.[1] |
| AV and VV timing | Routine AV/VV optimization is not required in all recipients; the SMART-AV trial found no benefit of echocardiographic or algorithmic AV-delay optimization over a fixed 120-ms AV delay in the general CRT population.[3] A pooled analysis of SMART-AV and SMART-CRT found improved CRT response with AV optimization in patients with interventricular delay ≥70 ms.[4] Consider electrocardiographic, echocardiographic, or device-algorithm–based optimization in patients with persistent symptoms, suboptimal hemodynamics, apparent non-response, or prolonged interventricular delay. |
| LV pacing vector and multipoint pacing | Quadripolar LV leads have lower rates of lead-related complications, lead deactivation, and lead dislodgement than bipolar leads and are associated with improved survival in meta-analysis.[5] Reprogramming a quadripolar LV vector may address phrenic nerve stimulation or high pacing thresholds. Multipoint pacing may be considered selectively, with attention to battery longevity and uncertain outcome benefit. |
Remote monitoring and heart-failure surveillance
Remote monitoring is standard care for patients with CRT devices. The remote-monitoring strategy should maintain connectivity and use individualized alerts for lead integrity, battery status, low BiV pacing, and clinically relevant atrial arrhythmias. Remote review of heart-failure diagnostics is reasonable when linked to a defined clinical response pathway.[6]
Minimum follow-up intervals include:
- Within 72 hours after implantation: in-person assessment.
- At 2–12 weeks after implantation: in-person assessment.
- Every 3–12 months for CRT-P and every 3–6 months for CRT-D: in-person or remote assessment.
- Annually until battery depletion: in-person assessment.
- Every 1–3 months when battery depletion is approaching: in-person or remote assessment.[7]
With consistent continuous connectivity and no recent alerts or active cardiac comorbidity requiring closer review, in-person visits may be extended to every 24 months (Class IIa, LOE B-R).[6]
Modern CRT devices may integrate activity, heart rate, respiratory rate, heart sounds, and impedance data. Multiparameter alerts can identify increased risk of heart-failure events and may support intervention within a defined clinical response pathway. A meta-analysis of six randomized trials found multiparameter-guided heart-failure management reduced the composite of all-cause death or heart-failure hospitalization (incidence rate ratio 0.83, 95% CI 0.71–0.99).[8] Thoracic impedance alone should not be used to manage congestion.[9]
Perform a follow-up echocardiogram approximately 3–12 months after implantation to assess LV remodeling, ejection fraction, ventricular volumes, and mitral regurgitation (Class I, LOE B-NR). Subsequent imaging should be directed by symptoms, device findings, or a change in clinical status.[9]
Management of apparent CRT non-response
Apparent non-response should prompt a structured reassessment rather than immediate device revision.
- Confirm true effective BiV pacing and exclude fusion, pseudofusion, intermittent LV capture, or low pacing delivery.
- Review device interrogation for lead thresholds, sensing, impedance trends, phrenic stimulation, arrhythmia burden, and programmed AV/VV timing.
- Identify and correct competing causes, including atrial fibrillation, frequent ventricular ectopy, progressive ischemia, valvular disease, and suboptimal guideline-directed medical therapy.
- Consider PVC-directed therapy, including catheter ablation, when PVC burden is high (commonly >10%–15%) and compromises effective BiV pacing.[1]
- Reassess LV lead position and myocardial substrate when clinically indicated.
- Refer persistent non-responders to a multidisciplinary heart-failure/electrophysiology service for consideration of optimization, arrhythmia treatment, lead revision, or alternative resynchronization strategies.[9]
In patients with AF, inadequate effective BiV pacing despite medical management should trigger reassessment of rate or rhythm control. Atrioventricular node ablation may improve pacing delivery in selected patients, but outcome evidence is not uniform across trials. The APAF-CRT trial showed a mortality benefit for AV-junction ablation plus de novo CRT implantation versus pharmacological rate control in patients with permanent AF and narrow QRS (≤110 ms).[10] In contrast, the CAAN-AF trial found no benefit of AV-node ablation versus medical rate control in patients with HFrEF, permanent AF, and pre-existing CRT-D.[11] These trials addressed different clinical questions and should not be directly compared. Decisions should be individualized according to pacing delivery, symptoms, ventricular function, procedural risk, and patient goals.
Generator replacement and device revision
At elective generator replacement, continue BiV pacing in patients with heart failure with improved ejection fraction or other evidence of clinical benefit from CRT. Evaluate lead performance, pacing thresholds, anticipated battery longevity, procedural risk, and the ongoing need for defibrillator therapy.[9]
In selected patients at elevated risk of CIED infection who undergo replacement or upgrade, consider evidence-based infection-prevention measures, including an absorbable antibiotic-eluting envelope where appropriate. In the WRAP-IT trial, the envelope reduced major CIED infections by 40% over 12 months in patients undergoing revision, replacement, upgrade, or de novo CRT-D implantation.[12][13]
Replacement of a CRT-defibrillator (CRT-D) with a CRT-pacemaker (CRT-P) should be individualized through shared decision-making. Consider prior appropriate ICD therapies, residual ventricular-arrhythmia risk, current LVEF, lead status, comorbidity, life expectancy, and patient preferences. In patients with LVEF recovery after primary-prevention CRT-D implantation, the 2025 appropriate-use criteria rate downgrade to CRT-P as “May Be Appropriate.”[14]
Device infection
Suspected cardiac implantable electronic device infection requires prompt electrophysiology and infectious-disease assessment. Definite pocket infection, lead infection, or CIED-related infective endocarditis generally requires complete system extraction plus culture-directed antimicrobial therapy; early extraction after diagnosis is associated with better outcomes.[13]
Empiric antibiotic therapy should include a vancomycin- or daptomycin-containing regimen because methicillin-resistant staphylococci are common causes of CIED infection; subsequent treatment should be culture-directed.[15] Recommended antimicrobial durations after extraction are 7 days for pocket erosion without purulence, 10 days for pocket infection with purulence, at least 2 weeks for bloodstream infection without valve involvement, and 4–6 weeks of parenteral therapy for valvular infective endocarditis.[13]
Reimplantation should occur only after infection control and negative blood cultures for at least 72 hours; in patients with valvular infective endocarditis, reimplantation should be delayed for at least 14 days after extraction. In selected patients with an ongoing indication for defibrillator protection, a wearable cardioverter-defibrillator may serve as a bridge to permanent device reimplantation. A contralateral implantation site is preferred when feasible. Long-term antimicrobial suppression without extraction is generally reserved for patients with prohibitive extraction risk or palliative goals.[13][16][17]
Magnetic resonance imaging
Before magnetic resonance imaging (MRI), verify the generator and every lead, including lead status and MRI-conditional labeling. An MRI-conditional system requires all components to meet the manufacturer’s conditions. Perform pre-scan interrogation, manufacturer-specific MRI programming, continuous physiologic monitoring during the scan, and post-scan interrogation.[18][19]
MRI in selected patients with non-conditional transvenous systems may be performed under an experienced institutional protocol after individualized risk assessment. Fractured, epicardial, or abandoned leads require particular caution and local protocol review.[18]
Goals of care and device deactivation
Discuss device management as part of advance-care planning at implantation and revisit it with clinical deterioration, recurrent hospitalization, transition to advanced heart-failure therapies, or palliative care. In CRT-D recipients, deactivation of ICD shock therapy may prevent distressing shocks near the end of life when consistent with informed patient preferences.[7]
Magnet application over a CRT-D generator can temporarily suspend tachyarrhythmia detection and therapy when formal reprogramming is not immediately available. Magnet response is programmable and may be disabled; Biotronik ICDs automatically re-enable tachyarrhythmia detection after 8 hours of continuous magnet application and, unlike other manufacturers, do not produce an audible tone when a magnet is applied. Confirm manufacturer- and programming-specific magnet behavior before relying on this approach. Pacing function is not affected by magnet application to an ICD/CRT-D.[20]
Do not automatically deactivate CRT pacing when ICD therapies are turned off. Withdrawal of BiV pacing can worsen heart-failure symptoms, particularly in pacing-dependent patients; decisions require individualized discussion, informed consent, and coordination among electrophysiology, heart-failure, and palliative-care teams.[21][22]
References
- ↑ 1.0 1.1 1.2 Madhavan M, Mulpuru SK, McLeod CJ, Cha YM, Friedman PA (2017). “Advances and Future Directions In Cardiac Pacemakers: Part 2 of a 2-Part Series”. Journal of the American College of Cardiology. 69 (2): 211–235. doi:10.1016/j.jacc.2016.10.064.
- ↑ Koplan BA, Kaplan AJ, Weiner S; et al. (2009). “Heart Failure Decompensation and All-Cause Mortality in Relation to Percent Biventricular Pacing in Patients With Heart Failure: Is a Goal of 100% Biventricular Pacing Necessary?”. Journal of the American College of Cardiology. 53 (4): 355–360. doi:10.1016/j.jacc.2008.09.043.
- ↑ Ellenbogen KA, Gold MR, Meyer TE; et al. (2010). “Primary Results From the SmartDelay Determined AV Optimization: A Comparison to Other AV Delay Methods Used in Cardiac Resynchronization Therapy (SMART-AV) Trial”. Circulation. 122 (25): 2660–2668. doi:10.1161/CIRCULATIONAHA.110.992552.
- ↑ Gold MR, Auricchio A, Leclercq C; et al. (2024). “Atrioventricular Optimization Improves Cardiac Resynchronization Response in Patients With Long Interventricular Electrical Delays: A Pooled Analysis of the SMART-AV and SMART-CRT Trials”. Heart Rhythm. 21 (9): 1686–1694. doi:10.1016/j.hrthm.2024.03.1783.
- ↑ Erath JW, Benz AP, Hohnloser SH, Vamos M (2019). “Clinical Outcomes After Implantation of Quadripolar Compared to Bipolar Left Ventricular Leads in Patients Undergoing Cardiac Resynchronization Therapy: A Systematic Review and Meta-Analysis”. Europace. 21 (10): 1543–1553. doi:10.1093/europace/euz190.
- ↑ 6.0 6.1 Ferrick AM, Raj SR, Deneke T; et al. (2023). “2023 HRS/EHRA/APHRS/LAHRS Expert Consensus Statement on Practical Management of the Remote Device Clinic”. Heart Rhythm. 20 (9): e92–e144. doi:10.1016/j.hrthm.2023.03.1525.
- ↑ 7.0 7.1 Epstein AE, DiMarco JP, Ellenbogen KA; et al. (2013). “2012 ACCF/AHA/HRS Focused Update Incorporated Into the ACCF/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities”. Journal of the American College of Cardiology. 61 (3): e6–e75. doi:10.1016/j.jacc.2012.11.007.
- ↑ Zito A, Restivo A, Ciliberti G; et al. (2023). “Heart Failure Management Guided by Remote Multiparameter Monitoring: A Meta-Analysis”. International Journal of Cardiology. 388: 131163. doi:10.1016/j.ijcard.2023.131163.
- ↑ 9.0 9.1 9.2 9.3 Chung MK, Patton KK, Lau CP; et al. (2023). “2023 HRS/APHRS/LAHRS Guideline on Cardiac Physiologic Pacing for the Avoidance and Mitigation of Heart Failure”. Heart Rhythm. 20 (9): e17–e91. doi:10.1016/j.hrthm.2023.03.1538.
- ↑ Brignole M, Pentimalli F, Palmisano P; et al. (2021). “AV Junction Ablation and Cardiac Resynchronization for Patients With Permanent Atrial Fibrillation and Narrow QRS: The APAF-CRT Mortality Trial”. European Heart Journal. 42 (46): 4731–4739. doi:10.1093/eurheartj/ehab569.
- ↑ Sanders P, Ariyaratnam JP, Stiles MK; et al. (2026). “Cardiac Resynchronization Therapy With or Without Atrioventricular Node Ablation in Atrial Fibrillation: The CAAN-AF Trial”. European Heart Journal. doi:10.1093/eurheartj/ehag206.
- ↑ Tarakji KG, Mittal S, Kennergren C; et al. (2019). “Antibacterial Envelope to Prevent Cardiac Implantable Device Infection”. New England Journal of Medicine. 380 (20): 1895–1905. doi:10.1056/NEJMoa1901111.
- ↑ 13.0 13.1 13.2 13.3 Baddour LM, Esquer Garrigos Z, Rizwan Sohail M; et al. (2024). “Update on Cardiovascular Implantable Electronic Device Infections and Their Prevention, Diagnosis, and Management: A Scientific Statement From the American Heart Association”. Circulation. 149 (2): e201–e216. doi:10.1161/CIR.0000000000001187.
- ↑ Russo AM, Desai MY, Do MM; et al. (2025). “ACC/AHA/ASE/HFSA/HRS/SCAI/SCCT/SCMR 2025 Appropriate Use Criteria for Implantable Cardioverter-Defibrillators, Cardiac Resynchronization Therapy, and Pacing”. Journal of the American College of Cardiology. 85 (11): 1213–1285. doi:10.1016/j.jacc.2024.11.023.
- ↑ Palmeri NO, Kramer DB, Karchmer AW, Zimetbaum PJ (2021). “A Review of Cardiac Implantable Electronic Device Infections for the Practicing Electrophysiologist”. JACC: Clinical Electrophysiology. 7 (6): 811–828. doi:10.1016/j.jacep.2020.10.020.
- ↑ Chesdachai S, Esquer Garrigos Z, DeSimone CV, DeSimone DC, Baddour LM (2024). “Infective Endocarditis Involving Implanted Cardiac Electronic Devices: JACC Focus Seminar 1/4”. Journal of the American College of Cardiology. 83 (14): 1326–1337. doi:10.1016/j.jacc.2023.11.036.
- ↑ Kusumoto FM, Schoenfeld MH, Wilkoff BL; et al. (2017). “2017 HRS Expert Consensus Statement on Cardiovascular Implantable Electronic Device Lead Management and Extraction”. Heart Rhythm. 14 (12): e503–e551. doi:10.1016/j.hrthm.2017.09.001.
- ↑ 18.0 18.1 Indik JH, Gimbel JR, Abe H; et al. (2017). “2017 HRS Expert Consensus Statement on Magnetic Resonance Imaging and Radiation Exposure in Patients With Cardiovascular Implantable Electronic Devices”. Heart Rhythm. 14 (7): e97–e153. doi:10.1016/j.hrthm.2017.04.025.
- ↑ American College of Radiology. American College of Radiology Manual on MR Safety: 2024 Update and Revisions. 2024.
- ↑ Wan EY, Rogers AJ, Lavelle M; et al. (2024). “Periprocedural Management and Multidisciplinary Care Pathways for Patients With Cardiac Implantable Electronic Devices: A Scientific Statement From the American Heart Association”. Circulation. 150 (8): e183–e196. doi:10.1161/CIR.0000000000001264.
- ↑ Graven LJ, Kitko L, Abshire Saylor M; et al. (2025). “Palliative Care and Advanced Cardiovascular Disease in Adults: Not Just End-of-Life Care: A Scientific Statement From the American Heart Association”. Circulation. 151 (21): e1030–e1042. doi:10.1161/CIR.0000000000001323.
- ↑ Kusumoto FM, Schoenfeld MH, Barrett C; et al. (2019). “2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay”. Journal of the American College of Cardiology. 74 (7): e51–e156. doi:10.1016/j.jacc.2018.10.044.
“`
Complications
Editor-In-Chief: C. Michael Gibson, M.S., M.D. [1]; Associate Editor(s)-In-Chief:: Bhaskar Purushottam, M.D. [2]
Overview
Bleeding, infection and lead dislodgement are not uncommon complications of cardiac resynchronization therapy.
Common Complications
- Bleeding and Hematomas: The incidence in clinical trials is 2.4%; in routine clinical practice, the actual incidence of pocket hematomas is probably higher as the trials only reported those hematomas, which needed surgical intervention. It is important to note that early re-intervention of pocket hematomas is associated with a 15-fold increase in infection[1].
- Infection
- Lead dislodgement. (CRT trials demonstrated a rate verying from 2.9% to 10.6%; the MIRACLE-ICD[2] study demonstrated a higher occurence of lead dislodgement with left ventricular lead than right atrial and right ventricular leads-6.8%, 15 and 0.6% respectively).
Rare Complications
- Pneumothorax (0.9% in CRT trials, Medicare registry[3] reported 1.2%),
- Myocardial injury
- Coronary sinus dissection (1.3%) or perforation (1.3%)(complication rate related to coronary veins has been reported in 2%)
- Pericardial tamponade.
- Pocket erosion
- Lead fracture
- Inappropriate phrenic nerve stimulation: Given the proximity of the posterior wall of the left ventricle to the phrenic nerve, there remains the risk of inappropriate phrenic nerve stimulation.
Mortality
For CRT patients, the average in-hospital mortality is 0.3% and the 30 day mortality is 0.7%. Reynolds et al.[3] showed a 1.1% in-hospital mortality in 30,984 Medicare patients undergoing CRT. This disparity in the in-hospital mortality rates can be explained by the strict inclusion criteria of the trials, where the selected patients are healthier than the unselected patients in clinical practice.[4]
References
- ↑ Klug D, Balde M, Pavin D, Hidden-Lucet F, Clementy J, Sadoul N; et al. (2007). “Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study”. Circulation. 116 (12): 1349–55. doi:10.1161/CIRCULATIONAHA.106.678664. PMID 17724263.
- ↑ Young JB, Abraham WT, Smith AL, Leon AR, Lieberman R, Wilkoff B; et al. (2003). “Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD Trial”. JAMA. 289 (20): 2685–94. doi:10.1001/jama.289.20.2685. PMID 12771115.
- ↑ 3.0 3.1 Reynolds MR, Cohen DJ, Kugelmass AD, Brown PP, Becker ER, Culler SD; et al. (2006). “The frequency and incremental cost of major complications among medicare beneficiaries receiving implantable cardioverter-defibrillators”. J Am Coll Cardiol. 47 (12): 2493–7. doi:10.1016/j.jacc.2006.02.049. PMC 1800827. PMID 16781379.
- ↑ van Rees JB, de Bie MK, Thijssen J, Borleffs CJ, Schalij MJ, van Erven L (2011). “Implantation-related complications of implantable cardioverter-defibrillators and cardiac resynchronization therapy devices a systematic review of randomized clinical trials”. J Am Coll Cardiol. 58 (10): 995–1000. doi:10.1016/j.jacc.2011.06.007. PMID 21867832.
Looking for the patient version?
© 2026 MyEClinic – IFTM Institut für Telematik in der Medizin GmbH
